SmartTrap: automated precision experiments with optical tweezers.

Selin M, Ciarlo A, Pesce G, Bengtsson L, Camunas-Soler J, Sundar Rajan V, … Volpe G
Optical tweezers

Results

ObservableValueSource sentence
other0.04 pNIn this way, the SmartTrap achieves a force resolution of 0.04 pN and a distance resolution of 1 nm when using the PSD sensors and about 10 nm when using video microscopy.
other480,000 charge groups per particleFrom the fit shown in Fig. 6, the number of charge groups was found to be Z ≈ 480,000.

Auto-extracted; verify against the paper before reuse.

Biological system

  • lambda-DNA EagI fragment F3 (11.8 kb) · dna
  • lambda-DNA EagI fragment F1 (19.9 kb), minor population · dna
  • Digoxigenin-labelled PCR handles (224 and 266 bases) · dna
  • Human red blood cell · cell

Data availability

Abstract

Optical tweezers are widely used in single-molecule biophysics, cell biomechanics and soft matter physics, but require a human operator, limiting throughput and repeatability. Here we present a smart optical tweezers platform, named SmartTrap, capable of performing complex experiments autonomously by integrating real-time three-dimensional particle tracking, custom electronics and a microfluidics system. Through a series of experiments, we demonstrate it can operate continuously, acquiring high-precision data over extended periods of time. By bridging the gap between manual experimentation and autonomous operation, SmartTrap establishes a robust and open-source framework for the next generation of optical tweezers research, capable of performing large-scale studies in single-molecule biophysics, cell mechanics and colloidal science with minimal experimental overhead and operator bias.