High-force NdFeB-based magnetic tweezers device optimized for microrheology experiments.
Results
| Observable | Value | Source sentence |
|---|---|---|
| Elastic modulus | 1 kPa | e employed to enable the measurement of materials with viscoelastic moduli as high as ∼1 kPa. |
Auto-extracted; verify against the paper before reuse.
Biological system
- Cytoplasm (microrheology) · cell
Data availability
Not checked (no full text)
Abstract
We present the design, calibration, and testing of a magnetic tweezers device that employs two pairs of permanent neodymium iron boron magnets surrounded by low-carbon steel focusing tips to apply large forces to soft materials for microrheology experiments. Our design enables the application of forces in the range of 1-1800 pN to ∼4.5 μm paramagnetic beads using magnet-bead separations in the range of 0.3-20 mm. This allows the use of standard coverslips and sample geometries. A high speed camera, custom LED-based illumination scheme, and mechanically stabilized measurement platform are employed to enable the measurement of materials with viscoelastic moduli as high as ∼1 kPa.