Interface shear microrheometer with an optically driven oscillating probe particle.

Park CY, Ou-Yang HD, Kim MW
The Review of scientific instruments 2011 doi:10.1063/1.3627410 PubMed
Optical tweezers

Biological system

Data availability

Not checked (no full text)

Abstract

We report the first experimental demonstration of an active interfacial shear microrheometer (ISMR) that uses a particle trapped by oscillating optical tweezers (OT) to probe the shear modulus G(s)(*)(ω) of a gas/liquid interface. The most significant advantages of the oscillating OT in a rheology study are: (1) very high sensitivity compared to other active microrheology methods and (2) the ability to measure both the real and imaginary components of the complex shear modulus without relying on the use of Kramers-Kronig relation, which can be problematic at low frequencies for most of the passive methods. We demonstrate the utilities of our ISMR in two case studies: (1) a 1,2-dipalmitoyl-sn-glycero-3-phosphocholine monolayer and (2) a composite of poly(styrene sulfonate) and dioctadecyldimethylammonium at the air/water interface in regimes where no other active instruments can explore.